MLE Fit with Global
Parameters Separates FAD and FMN in Metabolic FLIM Data
W. Becker, A. Bergmann, Becker &
Hickl GmbH
Berlin,
Germany
Abstract:
Metabolic FLIM is based on
the analysis of NADH and/or FAD fluorescence decay functions. The decay
functions change with the metabolic state of the cells. Double-exponential
decay analysis delivers the amplitudes of the fast decay components, a1, or the
ratios of the amplitudes of the fast and slow components, a1/a2. These values
are used as quantitative indicators of the metabolic state. The double-exponential
model is reasonably correct for the NADH data. However, it is not entirely
correct for FAD. Precision analysis of the FAD data shows that there is a third
decay component of about 4 ns to 5 ns and an amplitude of 5 % to
10 %. Therefore, triple-exponential decay analysis is required to obtain
quantitative metabolic-FLIM results from FAD. However, the photon numbers
available in metabolic-FLIM data are insufficient to derive reliable component
amplitudes by conventional point-by-point analysis. We therefore developed an
MLE-based 'Global Fit' procedure that exploits the fact that the component
lifetimes do not vary significantly within a population of cells of similar
type. The procedure delivers low-noise images of the component amplitudes,
representing the relative amounts of bound FAD, free FAD, and FMN.
Keywords: Metabolic FLIM, Global Fit, FAD, NADH
Principle of Metabolic FLIM
FAD is a natural fluorophore that is
present in almost any cell and any biological tissue. Together with NADH, it is
involved in the metabolism of the cell. FAD exists in a bound and an unbound
form. The two forms have different fluorescence lifetimes. The lifetime of the
bound form is on the order of 300 ps, the lifetime of the unbound form is
about 1 ns. The relative amounts of bound and unbound FAD change with the
type of the metabolism. Under glycolytic conditions, the amount of bound FAD decreases,
and the amount of free FAD increases. The effect shows up in the amplitudes of
the decay components. In the glycolytic state the amplitude of the fast
component, a1, decreases, and the amplitude of the slow component, a2, increases.
A similar effect exists for NADH. However, here the effect on the decay
function is opposite to that for FAD. The reason is that for NADH the fast
component comes from free NADH, and the slow one from bound NADH. Please see [1] for a detailed description.
In principle, metabolic FLIM can be
performed on NADH alone [1, 6]. However, neglecting the FAD would mean
discarding valuable information. The bh metabolic FLIM technique therefore aims
at recording FAD and NADH signals simultaneously. The problem of simultaneous
NADH and FAD FLIM is the overlap of the excitation and emission spectra. Clean
separation of NADH and FAD fluorescence can only be obtained if different
excitation and emission wavelengths are used [1]. The problem is not, as
commonly believed, to excite the FAD. It excites at almost any wavelength from
350 to 460 nm. The problem is to avoid NADH excitation when FAD is
excited. To do so, the bh metabolic FLIM technique uses two multiplexed lasers
with wavelengths of 370 nm and 405 nm, or 780 nm to 920 nm for
two-photon excitation [2]. Multidimensional TCSPC records the signals into separate
FLIM images [1].
FAD FLIM Data
Fig. 1 shows FAD FLIM data recorded with a
bh metabolic FLIM system. The data were analysed by bh's SPCImage NG using an
MLE (Maximum-Likelihood Estimation) procedure. MLE was used because it reaches
a significantly better fit accuracy for multi-exponential decays than the
traditional weighted least-square fit [4].
The cells marked on the left are normal
cells, the cells marked on the right are tumor cells. Decay functions at the
cursor position and component amplitudes are shown at the bottom. The cells
show the expected behaviour: The amplitude, a1, is high in the good cells, and
lower in the tumor cell. A similar behaviour is found also in other cells. This
shows that FAD FLIM can, in principle, be used to evaluate the metabolic state
of cells.

Fig. 1: FAD FLIM image of human epithelial bladder cells. The cells marked
on the left are normal cells, the cells marked on the right are tumor cells.
Decay functions at the cursor position and component amplitudes shown at the
bottom.
However, precision decay analysis of the
same data shows that the assumption of a purely double-exponential decay is not
entirely correct. Analysis of the combined decay data from larger areas is show
in Fig. 2. The combined data contain a much high number of photons than
single-pixel data and thus deliver higher fit accuracy. The results show that
there is a third decay component of about 4.5 ns lifetime. The amplitude
is a3 = 3.9 % in the good cells and a3 = 8.9 % in the
tumor cells.

Fig. 2: Decay analysis of the combined pixels from larger areas. Left:
Good cells. Right: Tumor cells. The decay data show that there is a third decay
component of about 4.5 ns lifetime. The amplitude is a3 = 3.9 % in
the good cells and a3 = 8.9 % in the tumor cells.
The most likely source of the third decay
component is the presence of FMN. FMN has a molecular structure largely similar
to FAD, a similar emission spectrum, and a fluorescence lifetime in the range
of 4 to 5 ns. FMN is not directly involved in the cell metabolism, and its
lifetime does not change with the metabolic state. The data shown above
therefore confirm the assumption that the third component is indeed FMN.
The presence of FMN does not change the
fact that FAD data can be used to evaluate the metabolic state. It does,
however, raise the question of whether double-exponential fit results of FAD can
be considered a quantitative expression of the state of the cells.
Triple-Exponential Analysis
Quantitative determination of the amounts
of bound and unbound FAD in presence of FMN requires triple-exponential decay
analysis. Triple-exponential analysis of FAD / FMN data delivers good results
as long as only a 'lifetime image', i.e. an image of the amplitude-weighted
decay time, tm, of the three decay components is considered. A
triple-exponential tm image of the data shown above is displayed in Fig. 3.

Fig. 3: tm Image of the sample shown in the figures above. MLE fit with
triple exponential model, tm is the amplitude-weighted average of the component
lifetimes. Right: Decay curve at the cursor position. Residuals of the fit are
shown under the decay curve. Note the perfectly flat residuals.
MLE analysis with the triple-exponential
model delivers an excellent fit quality, confirmed by the smooth residuals. However,
metabolic FLIM requires more than the buildup of a lifetime image - it requires
the extraction of the amplitudes of the decay components. An attempt to build
up a1, a2, and a3 images from the same data does not deliver a satisfactory
result, see Fig. 4.

Fig. 4: An
attempt to determine the amplitudes a1, a2, and a3 of the triple-exponential
fit result. Compare with tm image in Fig. 3. The amplitudes are obtained at a
much lower signal-to-noise ratio than the average lifetime, tm.
Global Fit
Experience shows that the component
lifetimes in FAD and NADH images do not vary significantly over the area of a
sample. The lifetimes may differ for different cell types, but within one and
the same cell type they are essentially constant. Changes in the apparent
lifetime are essentially caused by variation in the component amplitudes. This
opens the opportunity to fit the decay data with 'Global Parameters'. The
procedure is also known as a 'Global Fit' [4]. A global fit assumes that one or several
of the component lifetimes of a multi-exponential decay are constant over the entire
FLIM image. By using this as a-priori knowledge, the component amplitudes are
obtained at a significantly better accuracy than by a multi-exponential fit
with all decay parameters floating. Please note that a global fit is not the
same as a fit with fixed component lifetimes: A fit with fixed lifetimes [4]
uses component lifetimes defined by the user. These are not necessarily correct
for the particular sample. A global fit determines a combination of lifetimes
which yields the best fit for the sample under investigation. This avoids
subjective influence of the user on the results.
A suitable global fit procedure is
available in bh's SPCIMage NG, version 9.073 [4]. A fit with global component
lifetimes, t1, t2, t3, is shown in Fig. 5. As can be seen in Fig. 5, the global-fit
procedure reaches a massively increased signal-to-noise ratio of the component
amplitudes and, consequently, of the amounts of bound FAD, free FAD, and FMN.

Fig. 5: Amplitudes of bound FAD, a1, free FAD, a2, and FMN, a3, determined
by a fit with global parameters, t1, t2, and t3. MLE fit. The result displays a
massively increased signal-to-noise ratio of the component amplitudes, compare Fig.
4.
Concluding Remarks
The inclusion of FMN fluorescence in the
analysis by global fitting of FAD data is a big step towards quantitative
metabolic imaging of FAD. It not only supplements metabolic FLIM of NADH but
also opens an opportunity to record metabolic data at excitation wavelengths in
the visible range. Compared with UV excitation of NADH, the longer excitation
wavelength results in reduced aberrations in the optical system, lower
background fluorescence, and, probably, lower phototoxic effects to the cells.
A concern related to the analysis technique
may be the necessity of global-fit analysis. A global fit is enormously computation-intensive.
Processing times with standard CPU processing are therefore unacceptably long. So
far, the long processing times have prevented, or at least impeded, the
widespread introduction of global fitting. The situation has changed with the
availability of SPCImage NG. SPCImage NG runs the analysis on a GPU. In the
GPU, data of a large number of pixels are processed in parallel. GPU processing
reduces the calculation time of a global fit from previously hours to a few
minutes and less. Calculation time is therefore no longer a concern [3]. Decay analysis is further improved by MLE
fitting, which significantly improves the fit reliability of multi-exponential
FLIM data [4].
References
1.
W. Becker, The bh TCSPC Handbook. 11th ed.
Becker & Hickl GmbH (2026)
2.
Becker & Hickl GmbH, DCS-120 Confocal and
Multiphoton FLIM Systems, user handbook, 10th ed. (2026). Available on
www.becker-hickl.com
3.
W. Becker, A. Bergmann, Fast GPU-Based Global
Fit of TCSPC FLIM Data. Application note, available on www.becker-hickl.com
(2026)
4.
SPCImage Data Analysis, in W. Becker, The bh
TCSPC Handbook. 11th ed. Becker & Hickl GmbH (2026)
5.
W. Becker, L. Braun, DCS-120 FLIM System Detects
FMN in Live Cells, application note, available on www.becker-hickl.com (2023).