MLE Fit with Global Parameters Separates FAD and FMN in Metabolic FLIM Data

Metabolic FLIM is based on the analysis of NADH and/or FAD fluorescence decay functions. The decay functions change with the metabolic state of the cells. Double-exponential decay analysis delivers the amplitudes of the fast decay components, a1, or the ratios of the amplitudes of the fast and slow components, a1/a2. These values are used as quantitative indicators of the metabolic state. The double-exponential model is reasonably correct for the NADH data. However, it is not entirely correct for FAD. Precision analysis of the FAD data shows that there is a third decay component of about 4 ns to 5 ns and an amplitude of 5 % to 10 %. Therefore, triple-exponential decay analysis is required to obtain quantitative metabolic-FLIM results from FAD. However, the photon numbers available in metabolic-FLIM data are insufficient to derive reliable component amplitudes by conventional point-by-point analysis. We therefore developed an MLE-based ‘Global Fit’ procedure that exploits the fact that the component lifetimes do not vary significantly within a population of cells of similar type. The procedure delivers low-noise images of the component amplitudes, representing the relative amounts of bound FAD, free FAD, and FMN.

Download PDF

MLE Fit with Global Parameters Separates FAD and FMN in Metabolic FLIM Data

W. Becker, A. Bergmann, Becker & Hickl GmbH

Berlin, Germany

Abstract: Metabolic FLIM is based on the analysis of NADH and/or FAD fluorescence decay functions. The decay functions change with the metabolic state of the cells. Double-exponential decay analysis delivers the amplitudes of the fast decay components, a1, or the ratios of the amplitudes of the fast and slow components, a1/a2. These values are used as quantitative indicators of the metabolic state. The double-exponential model is reasonably correct for the NADH data. However, it is not entirely correct for FAD. Precision analysis of the FAD data shows that there is a third decay component of about 4 ns to 5 ns and an amplitude of 5 % to 10 %. Therefore, triple-exponential decay analysis is required to obtain quantitative metabolic-FLIM results from FAD. However, the photon numbers available in metabolic-FLIM data are insufficient to derive reliable component amplitudes by conventional point-by-point analysis. We therefore developed an MLE-based 'Global Fit' procedure that exploits the fact that the component lifetimes do not vary significantly within a population of cells of similar type. The procedure delivers low-noise images of the component amplitudes, representing the relative amounts of bound FAD, free FAD, and FMN.

Keywords: Metabolic FLIM, Global Fit, FAD, NADH

 

Principle of Metabolic FLIM

FAD is a natural fluorophore that is present in almost any cell and any biological tissue. Together with NADH, it is involved in the metabolism of the cell. FAD exists in a bound and an unbound form. The two forms have different fluorescence lifetimes. The lifetime of the bound form is on the order of 300 ps, the lifetime of the unbound form is about 1 ns. The relative amounts of bound and unbound FAD change with the type of the metabolism. Under glycolytic conditions, the amount of bound FAD decreases, and the amount of free FAD increases. The effect shows up in the amplitudes of the decay components. In the glycolytic state the amplitude of the fast component, a1, decreases, and the amplitude of the slow component, a2, increases. A similar effect exists for NADH. However, here the effect on the decay function is opposite to that for FAD. The reason is that for NADH the fast component comes from free NADH, and the slow one from bound NADH. Please see [1] for a detailed description.

In principle, metabolic FLIM can be performed on NADH alone [1, 6]. However, neglecting the FAD would mean discarding valuable information. The bh metabolic FLIM technique therefore aims at recording FAD and NADH signals simultaneously. The problem of simultaneous NADH and FAD FLIM is the overlap of the excitation and emission spectra. Clean separation of NADH and FAD fluorescence can only be obtained if different excitation and emission wavelengths are used [1]. The problem is not, as commonly believed, to excite the FAD. It excites at almost any wavelength from 350 to 460 nm.  The problem is to avoid NADH excitation when FAD is excited. To do so, the bh metabolic FLIM technique uses two multiplexed lasers with wavelengths of 370 nm and 405 nm, or 780 nm to 920 nm for two-photon excitation [2]. Multidimensional TCSPC records the signals into separate FLIM images [1].

FAD FLIM Data

Fig. 1 shows FAD FLIM data recorded with a bh metabolic FLIM system. The data were analysed by bh's SPCImage NG using an MLE (Maximum-Likelihood Estimation) procedure. MLE was used because it reaches a significantly better fit accuracy for multi-exponential decays than the traditional weighted least-square fit [4].

The cells marked on the left are normal cells, the cells marked on the right are tumor cells. Decay functions at the cursor position and component amplitudes are shown at the bottom. The cells show the expected behaviour: The amplitude, a1, is high in the good cells, and lower in the tumor cell. A similar behaviour is found also in other cells. This shows that FAD FLIM can, in principle, be used to evaluate the metabolic state of cells.

Fig. 1: FAD FLIM image of human epithelial bladder cells. The cells marked on the left are normal cells, the cells marked on the right are tumor cells. Decay functions at the cursor position and component amplitudes shown at the bottom.

However, precision decay analysis of the same data shows that the assumption of a purely double-exponential decay is not entirely correct. Analysis of the combined decay data from larger areas is show in Fig. 2. The combined data contain a much high number of photons than single-pixel data and thus deliver higher fit accuracy. The results show that there is a third decay component of about 4.5 ns lifetime. The amplitude is a3 = 3.9 % in the good cells and a3 = 8.9 % in the tumor cells.

   

Fig. 2: Decay analysis of the combined pixels from larger areas. Left: Good cells. Right: Tumor cells. The decay data show that there is a third decay component of about 4.5 ns lifetime. The amplitude is a3 = 3.9 % in the good cells and a3 = 8.9 % in the tumor cells.

The most likely source of the third decay component is the presence of FMN. FMN has a molecular structure largely similar to FAD, a similar emission spectrum, and a fluorescence lifetime in the range of 4 to 5 ns. FMN is not directly involved in the cell metabolism, and its lifetime does not change with the metabolic state. The data shown above therefore confirm the assumption that the third component is indeed FMN.

The presence of FMN does not change the fact that FAD data can be used to evaluate the metabolic state. It does, however, raise the question of whether double-exponential fit results of FAD can be considered a quantitative expression of the state of the cells.

 

Triple-Exponential Analysis

Quantitative determination of the amounts of bound and unbound FAD in presence of FMN requires triple-exponential decay analysis. Triple-exponential analysis of FAD / FMN data delivers good results as long as only a 'lifetime image', i.e. an image of the amplitude-weighted decay time, tm, of the three decay components is considered. A triple-exponential tm image of the data shown above is displayed in Fig. 3.

   

Fig. 3: tm Image of the sample shown in the figures above. MLE fit with triple exponential model, tm is the amplitude-weighted average of the component lifetimes. Right: Decay curve at the cursor position. Residuals of the fit are shown under the decay curve. Note the perfectly flat residuals.

MLE analysis with the triple-exponential model delivers an excellent fit quality, confirmed by the smooth residuals. However, metabolic FLIM requires more than the buildup of a lifetime image - it requires the extraction of the amplitudes of the decay components. An attempt to build up a1, a2, and a3 images from the same data does not deliver a satisfactory result, see Fig. 4.

    

Fig. 4: An attempt to determine the amplitudes a1, a2, and a3 of the triple-exponential fit result. Compare with tm image in Fig. 3. The amplitudes are obtained at a much lower signal-to-noise ratio than the average lifetime, tm.

Global Fit

Experience shows that the component lifetimes in FAD and NADH images do not vary significantly over the area of a sample. The lifetimes may differ for different cell types, but within one and the same cell type they are essentially constant. Changes in the apparent lifetime are essentially caused by variation in the component amplitudes. This opens the opportunity to fit the decay data with 'Global Parameters'. The procedure is also known as a 'Global Fit' [4]. A global fit assumes that one or several of the component lifetimes of a multi-exponential decay are constant over the entire FLIM image. By using this as a-priori knowledge, the component amplitudes are obtained at a significantly better accuracy than by a multi-exponential fit with all decay parameters floating. Please note that a global fit is not the same as a fit with fixed component lifetimes: A fit with fixed lifetimes [4] uses component lifetimes defined by the user. These are not necessarily correct for the particular sample. A global fit determines a combination of lifetimes which yields the best fit for the sample under investigation. This avoids subjective influence of the user on the results.

A suitable global fit procedure is available in bh's SPCIMage NG, version 9.073 [4]. A fit with global component lifetimes, t1, t2, t3, is shown in Fig. 5. As can be seen in Fig. 5, the global-fit procedure reaches a massively increased signal-to-noise ratio of the component amplitudes and, consequently, of the amounts of bound FAD, free FAD, and FMN.

    

Fig. 5: Amplitudes of bound FAD, a1, free FAD, a2, and FMN, a3, determined by a fit with global parameters, t1, t2, and t3. MLE fit. The result displays a massively increased signal-to-noise ratio of the component amplitudes, compare Fig. 4.

Concluding Remarks

The inclusion of FMN fluorescence in the analysis by global fitting of FAD data is a big step towards quantitative metabolic imaging of FAD. It not only supplements metabolic FLIM of NADH but also opens an opportunity to record metabolic data at excitation wavelengths in the visible range. Compared with UV excitation of NADH, the longer excitation wavelength results in reduced aberrations in the optical system, lower background fluorescence, and, probably, lower phototoxic effects to the cells.

A concern related to the analysis technique may be the necessity of global-fit analysis. A global fit is enormously computation-intensive. Processing times with standard CPU processing are therefore unacceptably long. So far, the long processing times have prevented, or at least impeded, the widespread introduction of global fitting. The situation has changed with the availability of SPCImage NG. SPCImage NG runs the analysis on a GPU. In the GPU, data of a large number of pixels are processed in parallel. GPU processing reduces the calculation time of a global fit from previously hours to a few minutes and less. Calculation time is therefore no longer a concern [3]. Decay analysis is further improved by MLE fitting, which significantly improves the fit reliability of multi-exponential FLIM data [4].

 

References

1.              W. Becker, The bh TCSPC Handbook. 11th ed. Becker & Hickl GmbH (2026)

2.              Becker & Hickl GmbH, DCS-120 Confocal and Multiphoton FLIM Systems, user handbook, 10th ed. (2026). Available on www.becker-hickl.com

3.              W. Becker, A. Bergmann, Fast GPU-Based Global Fit of TCSPC FLIM Data. Application note, available on www.becker-hickl.com (2026)

4.              SPCImage Data Analysis, in W. Becker, The bh TCSPC Handbook. 11th ed. Becker & Hickl GmbH (2026)

5.              W. Becker, L. Braun, DCS-120 FLIM System Detects FMN in Live Cells, application note, available on www.becker-hickl.com (2023).

6.              R. Suarez-Ibarrola, L. Braun, P. Fabian Pohlmann, W. Becker, A. Bergmann, C. Gratzke, A. Miernik, K. Wilhelm, Metabolic Imaging of Urothelial Carcinoma by Simultaneous Autofluorescence Lifetime Imaging (FLIM) of NAD(P)H and FAD. Clinical Genitourinary Cancer (2020)

 