- Fast TCSPC Instrument Response: <120 ps FWHM
- Internal Cooler: Low Dark Count Rate
- Internal GHz Preamplifier: High Output Amplitude
- Internal Overload Protection
- Internal High Voltage Generator
- Simple +12 V Power Supply
- Standard C-Mount Thread Connection
- Small Size: 40 x 40 x 120 mm
- Direct Interfacing to All bh TCSPC Systems
Specifications
|
PMCS-150 |
||||
Cathode Type |
-4 |
-110 |
-113 |
-210 |
-20 |
Wavelength Range / nm 1) |
185 - 870 |
230 - 700 |
185 - 700 |
230 - 700 |
300 - 900 |
Detector Quantum Efficiency 1) |
25 % @ 380 nm |
35 % @ 350 nm |
43 % @ 350 nm |
18 % @ 490 nm |
|
Dark Count Rate (with cooling enabled 2)) / s-1 |
40 |
20 |
20 |
40 |
500 |
IRF Width, 1 mm Spot / Full Cathode Area (FWHM, typ. Value) |
100 ps / 120 ps |
||||
Single Electron Response Width (FWHM, typ. Value) |
1.5 ns |
||||
Single Electron Response Amplitude |
-100 to -200 mV |
||||
Count Rate |
> 5 MHz (continuous), > 100 MHz (Peak, < 1 µs) |
||||
Internal Overload Protection |
Down-regulation of PMT voltage |
||||
Recovery from Overload |
Automatically, when light intensity has returned to normal |
||||
Overload Indicator |
LED |
||||
Overload Signal |
TTL / CMOS, open drain, active low |
||||
Signal Output |
|
||||
Output Polarity |
negative |
||||
Output Connector |
SMA |
||||
Output Impedance |
50 Ω |
||||
Power Supply |
|||||
Power Supply |
+12 V, 300 mA 3) |
||||
General |
|||||
Cathode Diameter |
8 mm |
||||
Dimensions (Width x Height x Depth) |
(40 x 40 x 120) mm |
||||
Optical Adapter |
C-Mount female |
||||
Fibre Coupling |
SMA 905 or FC on request |
(1) According to Hamamatsu specifications
(2) Detector mounted on heat sink, temperature of detector base < 25 °C
(3) 100 mA when cooler is disabled