Specifications
|
PMC-150 |
||||
Cathode Type |
-4 |
-110 |
-113 |
-210 |
-20 |
Wavelength Range / nm 1) |
185 - 870 |
230 - 700 |
185 - 700 |
230 - 700 |
300 - 900 |
Detector Quantum Efficiency 1) |
25 % @ 380 nm |
35 % @ 350 nm |
43 % @ 350 nm |
18 % @ 490 nm |
|
|
Spectral sensitivity: click here | ||||
Dark Count Rate (typ. Value)/ s-1, Tcase = 22 °C |
20 |
5 |
5 |
10 |
200 |
Transit Time Spread (TCSPC IRF / FWHM, typ. Value) |
130 ps |
140 ps |
|||
Count Rate |
> 5 MHz (continuous), > 100 MHz (Peak, < 1 µs) |
||||
Single Electron Response Width (FWHM, typ. Value) |
1.5 ns |
||||
Single Electron Response Amplitude (Vgain = 0.9 V to 1 V) |
100 to 200 mV |
||||
Overload Indicator |
LED & Acoustic Signal |
||||
Overload Signal |
TTL / CMOS, active low |
||||
Signal Output |
|
||||
Output Polarity |
negative |
||||
Output Connector |
SMA |
||||
Output Impedance |
50 Ω |
||||
Power Supply |
|||||
(from DCC-100 Card) |
+12 V, -12 V (fan only), Cooler current 0.5 to 1 A |
||||
General |
|||||
Cathode Diameter |
8 mm |
||||
Dimensions (Width x Height x Depth) |
(76 x 111 x 56) mm |
||||
Optical Adapter |
C-Mount female |
||||
Fibre Coupling |
SMA 905 or FC on request |
(1) According to Hamamatsu specifications
Downloads
Documents
Datasheets
Application Notes
- Setting up High Gain Detector Electronics for TCSPC
- Detectors for High-Speed Photon Counting
- Two-Photon Fluorescence Excitation by Picosecond Diode Lasers
- Recording the Kautsky Effect by Fluorescence Lifetime Detection
- Timing Stability of TCSPC Experiments
- bh FLIM Systems for Nikon C1 Scanners
- Non-Descanned FLIM Systems for Olympus FV-1000 and FV-300 Multiphoton Microscopes
- Recording Z Scans with the DCS-120 Confocal Scanning FLIM System
- Microsecond Decay FLIM: Combined Fluorescence and Phosphorescence Lifetime Imaging
- TCSPC Fibre-Probe System with an Exchangeable Tip
- The PZ-FLIM-110 Piezo-Scanning FLIM System
- TCSPC System Records FLIM of a Rotating Object
- How (and why not) to Amplify PMT Signals