Fast-Acquisition TCSPC FLIM: What are the Options?
Wolfgang
Becker, Stefan Smietana, Becker & Hickl GmbH, Berlin, Germany
Abstract: This application note reviews the options
of fast-acquisition TCSPC FLIM. We describe the parameters which limit the
acquisition speed, both in the sample and in the recording system. We show that
a standard bh FLIM system is able to record and display FLIM images at a rate
much higher rate than commonly believed. We describe how faster acquisition
times and faster image rates are obtained in the bh FASTAC FLIM system.
Moreover, we show how fast dynamic changes in the fluorescence behaviour of a
sample can be recorded under conditions where the sample does not deliver the
count rate required for fast-acquisition FLIM.
General Considerations on Fast FLIM Acquisition
There is currently a run towards faster and
faster acquisition in fluorescence lifetime imaging (FLIM). Unfortunately, these
developments usually ignore the fact that, in most instances, the limitations
are in the sample rather than in the recording electronics. The fastest
recording system cannot record fast if the sample is not able to deliver the
photons needed to acquire the data at this speed. Moreover, recording speed is
usually obtained on the expense of time resolution or data complexity, such the
capability to record a fully resolved fluorescence decay function in each
pixel. These are exactly the features needed in typical FLIM applications, such
as metabolic imaging, protein interaction experiments, and other molecular
imaging applications. Ignoring these needs by just recording a lifetime in
the pixels of the image means reducing the fluorescence decay function to a
simple contrast parameter. In other words, Fast FLIM sacrifices the most
important features of FLIM for a single parameter (recording speed) which,
ironically, cannot be exploited in the typical applications. This application
note describes the parameters that determine the speed of FLIM recording, and
shows the way to record FLIM at short acquisition time without compromises in
time resolution or data complexity.
What Limits the Acquisition Speed of FLIM?
Signal-to-noise ratio
The signal-to-noise ratio (SNR) of FLIM
depends on the number of photons per pixel. Ideally, the SNR of the fluorescence
lifetime, t, is
with N = number of photons per pixel. For a
given photon rate obtained from the sample the number of photons, N, decreases
with decreasing acquisition time, and so does the signal-to-noise ratio of the
lifetime data.
The sample must feed the recording system with enough photons
The fastest FLIM system does not yield
images within a short acquisition time if the sample does not deliver the
required photon rate. For example, to record a 256 x 256 pixel image with an
SNR of 10 (or 10% standard deviation) 6.5 million photons are required. To
record the data within 1 second the photon rate must be 6.5×106 s-1.
To record a 512 x 512 pixel image at the same accuracy and
acquisition time a photon rate of 25×106 s‑1 would
be required. Photon rates this high can only be obtained from strongly stained
samples, such as the often used convallaria sample (stained with acridin
orange) or mouse kidney samples (stained with Alexa dyes). Samples used in FLIM
experiments normally contain far lower fluorophore concentrations and thus
deliver lower count rates. NADH in live cells yields no more than about 200,000
photons per second, and fluorescent proteins in FRET experiments no more than
500,000. Simultaneously, the requirements to the SNR are higher - lifetime
variations on the order of 2% have to be detected, and multi-exponential decay
analysis must be performed. Under these conditions, the required photon numbers
cannot be obtained within one or two seconds.
Photobleaching and photodamage
Attempts to increase the count rate by
increasing the excitation power induce lifetime changes in the sample, impair
the viability of the cells, or destroy the sample altogether. Fig. 1 shows two
examples. The left image shows a convallaria sample, the central region of
which has been scanned with a 405-nm laser for 1 minute at a laser power of
0.5 mW. The right image shows an NADH image of live cells. It was obtained
with two-photon excitation at 5% of the available laser power. The average
count rate over the entire image was 350,000 s-1. Within 20 seconds,
the irradiation caused damage in form of bright spots of extremely short
lifetime.
Fig. 1: Left: Convallaria image, region in the centre scanned with
405 nm laser. Right: Two-photon NADH image of live cells. In the bright
red spots photodamage has occurred, revealing itself by an ultra-fast decay
component.
It should be noted that photobleaching not
only has an impact on the imaging process, it also produces radicals. Radicals
cause photochemical stress to the cells. Attempts to reduce photobleaching by
increasing the fluorophore concentration do not reduce the amount of radicals
produced. The absolute amount of converted fluorophore remains the same, and so
does the photochemical stress to the cells.
Assuming we get enough photons:
How fast can TCSPC FLIM record?
TCSPC FLIM has a number of fascinating
features. It delivers an excellent time resolution [8, 9] and a near-ideal
photon efficiency [1, 2], it has multi-exponential decay-recording capability,
is able to record complex data in a multi-parameter space [1, 2, 3], and
simultaneously records FLIM and PLIM [6, 7]. These features are the basis of
metabolic FLIM, FLIM-FRET, multi-parameter and multi-wavelength FLIM, and
functional FLIM [3]. Should we sacrifice these functions and applications for
just faster acquisition - a feature which can be used only for a minority of
samples and a minority of experiments? No. Instead, we should find a way o
record at high photon rates while maintaining the superior functionality of
TCSPC FLIM.
How fast is a normal TCSPC FLIM system?
The Pile-Up Effect
As the speed-limiting effect of TCSPC FLIM
usually the Pile-Up is considered. Pile-up is the possible detection (and
loss) of a second photon in the same excitation pulse period with a previous
one [1, 2]. The probability of pile-up increases with increasing count rate. It
causes a distortion of the recorded decay curves, and a shift of the measured
decay times towards lower values, see Fig. 2.
Fig. 2: The pile-up effect. A second photon in the same pulse period with
a previous one is lost. The result is a distortion of the measured decay
profile.
It is commonly believed that the photon
rate must be lower than 0.1% of the pulse repetition rate, frep, to
avoid pile-up distortion. This is wrong. The Pile-Up Limit of 0.1% probably
stems from a typo in the early TCSPC literature. Correct is that the photon
rate can be as high as 10% (or 0.1 × frep)
without causing more than 5% of error in the recorded lifetimes [1, 2]. This is
100 times more than commonly believed!
Consequently, a standard TCSPC FLIM system can
record considerably faster than most users expect. Depending on the
expectations to the lifetime accuracy and on the number of pixels of the image acquisition
times between a few seconds and a few 100 ms can be achieved [10]. Fig. 3
shows two examples. The left image (256 x 256 pixels) was obtained
within 0.2 seconds, the right image (512 x 512 pixels) within 2
seconds.
Fig. 3: Fast acquisition with standard SPC-150 FLIM system. Left:
256 x 256 pixels, acquisition time 0.2 seconds. Right:
512 x 512 pixels, acquisition time 2 seconds. DCS-120 FLIM system,
online-lifetime display by SPCM software.
The pile-up free photon rate can be
increased by using several detectors and TCSPC modules in parallel and
combining the signals of the modules. Parallel-channel FLIM has been
demonstrated for two channels [19], four channels [18], and 8 channels [17]. Most
bh FLIM system already have two or four detectors and TCSPC channels [1, 14, 15,
16]. These parallel channels can easily by used to run fast FLIM on these
systems. Online combination of the channels has been implemented in SPCM,
version 9.78 [13].
The bh FASTAC FLIM System
The bh FASTAC FLIM system uses one detector
and four TCSPC modules. The photon pulses from the detector are distributed
into the TCSPC modules by a module called Photon Spinner [11]. The principle is shown in Fig. 4. As
usual, the photon pulses from the detector are passing a constant-fraction
discriminator, CFD. The output pulses of the CFD have a constant width and a
time independent of the pulse amplitude of the detector pulses [1]. The pulses from the CFD control a
four-way switch that distributes the pulses to four TCSPC modules. Every photon
pulse from the CFD rotates the switch by one position. The trick of the
solution is that every photon sets the switch not for the next photon, but for
itself. To do so, the photon pulses pass a delay line. Every photon arrives at
the switch a short time after the switching action has been completed. This
way, the Photon Spinner avoids the influence of possible switching transients
on the timing: Because every photon pulse arrives at the switch at a fixed time
after the switching action the sum of the photon pulse and the switching
transient is the same for all photons. It is also independent of the time of
the photon in the laser pulse period. Therefore, the time resolution (IRF
width) of the system and the differential nonlinearity of a bh FASTAC system is
the same as for a single-module system [11].
Fig. 4:
Principle of the bh FASTAC FLIM system
Fig. 5: FLIM recorded by the bh FASTAC
system. Left: 512 x 512 pixel image, recorded within 0.5 seconds.
Right: Decay function from a 10x10 pixel area in the centre of the image.
DCS-120 confocal FLIM system. Image and decay curve displayed by online FLIM
functions of SPCM.
Faster than Fast FLIM -
Temporal Mosaic Recording with Triggered Accumulation
What if the sample does not deliver the
photons for fast FLIM? In that case a fast FLIM system - no matter whether it
is a parallel-detector system or a FASTAC system - will deliver a similar image
as a standard single-channel system. Can we detect fast changes in the
fluorescence decay behaviour of a sample under these conditions? Surprisingly,
the answer is yes.
The way to do so is Temporal Mosaic FLIM.
Consider a sample in which, by some external stimulation, a lifetime change is
induced. Assume that the FLIM system records a series of FLIM images into
subsequent elements of a large data array (a mosaic of FLIM data) [1]. If the
recording starts with the stimulation, the result will be a fast time-series of
FLIM recordings. If each image is recorded in just one frame the series will be
as fast as the scanner can go.
Now, lets further assume that the
stimulation is applied to the sample periodically. What will happen? With every
stimulation the recording will run through all elements of the data array, and
the data will be accumulated. This is the idea of Temporal Mosaic FLIM with
Triggered Accumulation [1, 3, 5]. The result is a fast time series the
signal-to-noise ratio of which does no longer depend on the speed of the
series. For a given photon rate, it only depends on the total acquisition
time.
The principle is illustrated in Fig. 6. In
fact, the FLIM system records a photon distribution over the scan coordinates,
the times of the photons after the excitation pulses, and the times of the
photons after the stimulation.
Fig. 6:
Principle of Temporal Mosaic FLIM with Triggered Accumulation
A FLIM recording of the Calcium2+
transient in live neurons is shown in Fig. 7. The data were recorded with a
Zeiss LSM 7 multiphoton microscope in combination with a bh Simple Tau 150
(SPC-150) FLIM system [1, 5]. The time per image element was 38 ms.
Fig. 7: Calcium2+ transient in live neurons. Zeiss LSM 7MP,
SPC-150 FLIM module. Time per image element 38 ms, 100 stimulation periods
accumulated.
Resolving Transient Effects by Line-Scanning TCSPC
In a temporal mosaic recording as the one
shown in Fig. 7 the speed is limited by the minimum frame time of the scanner.
Faster sequences can be obtained by resonance scanners or by line scanning. A
typical galvanometer scanner can scan one line in about 1 to 2 milliseconds.
The FLIM system then builds up a photon distribution over the distance along
the line, the times of the photons after the excitation pulses, and the time
after the stimulation [4, 5]. An example is shown in Fig. 8. The time
resolution within the stimulation period is 2 ms.
Fig. 8: Recording transient lifetime effects by line scanning TCSPC. Left:
FLIM image of the sample, with selection of line scan within the sample. Right:
Line-scanning data, horizontal distance along the line, vertical time after
stimulation, colour fluorescence lifetime of Ca2+ sensor.
Summary
The recording speed of TCSPC FLIM is by one
part limited by the capability of the sample to deliver high photon rates, by
another part by the capability of the FLIM system to record this rate without
loss of photons and without distortion of the recorded decay profiles. The limiting
effect of the sample on the acquisition speed is usually underestimated,
whereas the limitations in the recording system are over-estimated. Samples for
molecular imaging experiments - the most widespread application of FLIM - do
not deliver the count rates required for fast-acquisition FLIM. Of the
recording side, TCSPC FLIM can record images surprisingly fast. In particular,
the much-feared pile-up effect is far less a speed limitation than commonly
believed.
As a consequence, standard TCSPC FLIM
systems are well capable of recording at the maximum photon rate typical FLIM
samples deliver. Nevertheless, there are samples which can deliver photon rates
beyond the capabilities of a normal TCSPC FLIM system. To exploit the count
rates from these samples for increasing the recording speed two solutions are
possible.
The signals can be split into several
channels optically, and recorded by several TCSPC modules in parallel. The
solution is simple and can be used in any bh FLIM system that has several
parallel TCSPC channels. All that is needed is a beamsplitter for the optical
signals to the detectors.
Fast acquisition with a single detector is
achieved by the bh FASTAC fast-acquisition FLIM system. The FASTAC system distributes
the photon pulses of one detector into four TCSPC channels electronically.
These fast FLIM principles achieve short
acquisition times without compromises in time resolution (IRF width), number of
time channels, or multi-dimensional recording capabilities of TCSPC FLIM. They
do, however, require a sample that deliver extremely high count rates. For
samples which do not deliver the required count rates a fast FLIM system is no
faster than a standard single-channel system.
Recording of fast changes in the
fluorescence behaviour under sample-limited (low-count rate) conditions can be
achieved by temporal mosaic FLIM with triggered accumulation. Different than parallel-channel
or FASTAC FLIM this technique does not require high count rates from the sample.
References
1.
W. Becker, The bh TCSPC handbook. Becker &
Hickl GmbH, 9th ed. (2021). Available on
www.becker-hickl.com
2.
W. Becker, Advanced time-correlated single
photon counting techniques. Springer, Berlin, Heidelberg, New York (2005)
3.
W. Becker (ed.), Advanced time-correlated single
photon counting applications. Springer, Berlin, Heidelberg, New York (2015)
4.
W. Becker, V. Shcheslavkiy, S. Frere, I.
Slutsky, Spatially Resolved Recording of Transient Fluorescence-Lifetime
Effects by Line-Scanning TCSPC. Microsc. Res. Techn. 77, 216-224 (2014)
5.
W. Becker, S. Frere, I. Slutsky, Recording Ca++ Transients
in Neurons by TCSPC FLIM. In:F.-J. Kao, G. Keiser, A. Gogoi, (eds.), Advanced
optical methods of brain imaging. Springer (2019)
6.
Becker & Hickl GmbH, Simultaneous
Phosphorescence and Fluorescence Lifetime Imaging by Multi-Dimensional TCSPC
and Multi-Pulse Excitation. Application note, available on www.becker-hickl.com
7.
W. Becker, V. Shcheslavskiy, A. Rück,
Simultaneous phosphorescence and fluorescence lifetime imaging by
multi-dimensional TCSPC and multi-pulse excitation. In: R. I. Dmitriev (ed.),
Multi-parameteric live cell microscopy of 3D tissue models. Springer (2017)
8.
Becker & Hickl GmbH, Sub-20ps IRF Width from
Hybrid Detectors and MCP-PMTs. Application note, available on
www.becker-hickl.com
9.
Becker & Hickl GmbH, Ultra-fast HPM
Detectors Improve NAD(P)H FLIM. Application note, available on
www.becker-hickl.com
10.
Becker & Hickl GmbH, SPCM Software Runs
Online-FLIM at 10 Images per Second. Application note, available on
www.becker-hickl.com
11.
Becker & Hickl GmbH, Fast-Acquisition TCSPC
FLIM System with sub-25 ps IRF Width. Application note, available on
www.becker-hickl.com
12.
Becker & Hickl GmbH, Fast-Acquisition
Multiphoton FLIM with the Zeiss LSM 880 NLO. Application note, available on
www.becker-hickl.com
13.
Becker & Hickl GmbH, New SPCM Version 9.80
Comes With New Software Functions. Application note, available on
www.becker-hickl.com
14.
Becker & Hickl GmbH, DCS-120 Confocal
Scanning FLIM Systems. User handbook, 7th ed. (2017). Available on
www.becker-hickl.com
15.
Becker & Hickl GmbH, FLIM Systems for
Zeiss LSM 710 / 780 / 880 Family Laser Scanning Microscopes. User Handbook, 7th
ed. (2017). Available on www.becker-hickl.com
16.
Becker & Hickl GmbH, FLIM systems for
Laser Scanning Microscopes. Overview brochure (2017), available on
www.becker-hickl.com.
17.
Becker & Hickl GmbH, An 8-Channel Parallel
Multispectral TCSPC FLIM System. Application note, available on
www.becker-hickl.com
18.
W. Becker, A. Bergmann, G. Biscotti, K. Koenig,
I. Riemann, L. Kelbauskas, C. Biskup, High-speed FLIM data acquisition by
time-correlated single photon counting, Proc. SPIE 5323, 27-35 (2004)
19.
V. Katsoulidou, A. Bergmann, W. Becker, How fast
can TCSPC FLIM be made? Proc. SPIE 6771, 67710B-1 to 67710B-7 (2007)
Contact:
Wolfgang Becker
Becker & Hickl GmbH
Berlin, Germany
Email: becker@becker-hickl.com
info@becker-hickl.com